Changchun
New Industries Optoelectronics Technology Co., Ltd.
Changchun Institute of Optics,
Fine Mechanics and
Physics, Chinese
Academy of Science
|
      
|
Products->Laser Power Meter
LASER POWER METER
|
The laser power meter manufactured by CNI is an optoelectronic device. The
main principle is to attenuate the
energy by colored optical glass and to receive the residual energy by
photoelectric cell uniformly. The power meter has
been tested for many parameters, such as:
linearity, repeatability, uniformity,
long term
stability and so on. The measurement
range of this power meter is from 0.01
milliwatt to 8 watts.
| Specifications: single
wavelength, high accuracy |
 |
| detector outside, convenient to use |
| good linearity,
good repeatability, good
uniformity |
|
measured wavelength (nm): 457、473、532、593.5、635、660、671、808、980、1064 |
Desk-top Power Meter
|
|
PM-XXX-100
|
PM-XXX-500
|
PM-XXX-3000 |
PM-XXX-8000 |
Specifications:
|
Parameter |
Laser Power
Meter Model |
|
PM-XXX-100
|
PM-XXX-500
|
PM-XXX-3000 |
PM-XXX-8000 |
| Power
Range |
≤100 mW |
≤500 mW |
≤3000 mW |
≤8000 mW |
| Power
Uncertainty |
±5% |
|
Precision |
0.01 mW |
0.1 mW |
1 mW |
10 mW |
| Meter
Head Mode |
Si |
|
Photosensitive Size |
φ25 mm |
| Temperature |
20±5 ℃
|
| Meter
Head Size |
Dia*Length=39*62(mm)
|
Dia*Length=43*74(mm)
|
Dia*Length=65*102(mm)
|
| Meter
Box Size |
Length*Width*Height=86*47*96(mm) |
| Weight |
950 g |
| Power |
AC 85~260 V |
Portable Power Meter
Capability
Specification:
|
Parameter |
Laser
Power Meter Model |
|
PM-XXX-100 |
PM-XXX-500 |
PM-XXX-3000 |
PM-XXX-5000 |
| Power
Range |
≤100 mW |
≤500 mW |
≤3000
mW |
≤5000 mW |
| Power
Uncertainty |
±5% |
| Precision |
0.01 mW |
0.1 mW |
0.1 mW
|
1 mW |
| Meter Head
Mode |
Si |
|
Photosensitive Size |
φ25 mm |
| Temperature |
20±5 ℃ |
| Meter Head Size |
Dial*Length=35*62(mm) |
Dial*Length=43*74(mm) |
Dial*Length=65*102(mm) |
| Meter Box
Size |
Length*Width*Height=145*80*36.5 mm |
| LED Show
Size |
50*20 mm
|
| Weight |
700 g |
| Power |
9 V battery
, 6F22 |
Order
Information:
PM-XXX-500:
PM: refer
to Power Meter, XXX: refer to measured wavelength, 500: refer
to the maximum range is 500 milliwatts.
For example: If the measured
wavelength is 532 nm and the maximum range of this detector is 3 W, the
according
model is PM-532-3000.
Notes:
1. When the power meter is
manufactured, it only can be used for the specified single wavelength.
2. Over range using is forbidden.
3. The admirable incidence angle of the laser on the surface of the detector
is zero.
4. The calibration date of this power meter is written on the qualified cards. It need
to be calibrated for every year.
5. The power meter is warranted for one year.
|